BACKGROUND: The d-ROMs test for the evaluation of serum hydroperoxides (HP) is simple, reliable, and cheap. Furthermore, it can easily be adapted to automated analyzers. Changing from the manual to an automated procedure allows the simultaneous processing of a large number of samples in a greatly reduced time, avoiding manual handling of samples and reagents and reducing variability sources. METHODS: This study was performed to adjust the manual procedure to a routine automated method in the clinical laboratory. We carried out the d-ROMs test in sera from 90 subjects of both sexes (34 men and 56 women) with age ranging from 20 to 80 years (mean 51+/-14 years). All subjects were free from acute or chronic inflammatory disease, immunological disease and history or evidence of malignancy. Subjects were not on vitamin and/or antioxidant therapies. RESULTS: The detection limit of the assay was 40 AU. Linearity was observed up to 475 AU. The recovery ranged between 97% and 105%. Within- and between-run imprecision was 5%. The mean HP value was 304+/-8 AU, with no significant difference between men (291+/-10 AU) and women (311+/-11 AU). A significant positive correlation was observed between age and HP in the whole population (r
Automation and validation of a fast method for the assessment of in vivo oxidative stress levels
Vassalle C;Ndreu R;
2006
Abstract
BACKGROUND: The d-ROMs test for the evaluation of serum hydroperoxides (HP) is simple, reliable, and cheap. Furthermore, it can easily be adapted to automated analyzers. Changing from the manual to an automated procedure allows the simultaneous processing of a large number of samples in a greatly reduced time, avoiding manual handling of samples and reagents and reducing variability sources. METHODS: This study was performed to adjust the manual procedure to a routine automated method in the clinical laboratory. We carried out the d-ROMs test in sera from 90 subjects of both sexes (34 men and 56 women) with age ranging from 20 to 80 years (mean 51+/-14 years). All subjects were free from acute or chronic inflammatory disease, immunological disease and history or evidence of malignancy. Subjects were not on vitamin and/or antioxidant therapies. RESULTS: The detection limit of the assay was 40 AU. Linearity was observed up to 475 AU. The recovery ranged between 97% and 105%. Within- and between-run imprecision was 5%. The mean HP value was 304+/-8 AU, with no significant difference between men (291+/-10 AU) and women (311+/-11 AU). A significant positive correlation was observed between age and HP in the whole population (rI documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.