A velocity map imaging/ion time-of-flight spectrometer designed specifically for pump-probe experiments combining synchrotron and laser radiations is described. The in-house built delay line detector can be used in two modes: the high spatial resolution mode and the coincidence mode. In the high spatial resolution mode a kinetic energy resolution of 6% has been achieved. The coincidence mode can be used to improve signal-to-noise ratio for the pump-probe experiments either by using a gate to count electrons only when the laser is present or by recording coincidences with the ion formed in the ionization process.
A photoelectron velocity map imaging spectrometer for experiments combining synchrotron and laser radiations
P Bolognesi;M Coreno;L Avaldi
2011
Abstract
A velocity map imaging/ion time-of-flight spectrometer designed specifically for pump-probe experiments combining synchrotron and laser radiations is described. The in-house built delay line detector can be used in two modes: the high spatial resolution mode and the coincidence mode. In the high spatial resolution mode a kinetic energy resolution of 6% has been achieved. The coincidence mode can be used to improve signal-to-noise ratio for the pump-probe experiments either by using a gate to count electrons only when the laser is present or by recording coincidences with the ion formed in the ionization process.File in questo prodotto:
Non ci sono file associati a questo prodotto.
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.