The design of a novel therapeutic drug monitoring (TDM) point-of-care-testing (POCT) biochip for immunosuppressants detection in transplanted patients is described. The chip consists of two polymeric parts, a top PMMA slide and a bottom ZEONOR® thin foil, bonded together by means of a pressure sensitive adhesive tape. The tape, with lower refractive index, is shaped in order to obtain a microfluidic multi-channel array. The optical signal, coming from an external light source and travelling along the ZEONOR® thin foil, excites the fluorescent sensing layer immobilized onto the fluidic channels. Preliminary tests with the bioassay implementation for tacrolimus detection are reported. © 2014 SPIE.

A newly designed optical biochip for a TDM-POCT device

Trono Cosimo;Giannetti Ambra;Tombelli Sara;Bernini Romeo;Persichetti Gianluca;Testa Genni;Baldini Francesco
2014

Abstract

The design of a novel therapeutic drug monitoring (TDM) point-of-care-testing (POCT) biochip for immunosuppressants detection in transplanted patients is described. The chip consists of two polymeric parts, a top PMMA slide and a bottom ZEONOR® thin foil, bonded together by means of a pressure sensitive adhesive tape. The tape, with lower refractive index, is shaped in order to obtain a microfluidic multi-channel array. The optical signal, coming from an external light source and travelling along the ZEONOR® thin foil, excites the fluorescent sensing layer immobilized onto the fluidic channels. Preliminary tests with the bioassay implementation for tacrolimus detection are reported. © 2014 SPIE.
2014
Istituto di Fisica Applicata - IFAC
Istituto per il Rilevamento Elettromagnetico dell'Ambiente - IREA
Inglese
photonic west
8976
9780819498892
http://www.scopus.com/record/display.url?eid=2-s2.0-84900530396&origin=inward
Fluorescence anisotropy
Immunosuppressants
Microfluidic multichannel array
Optical biochip
Point-of-care-testing
Therapeutic drug monitoring
14
none
Berrettoni, C; Trono, Cosimo; Berneschi, Simone; Giannetti, Ambra; Tombelli, Sara; Bernini, Romeo; Grimaldi Angelica, Immacolata; Persichetti, Gianluc...espandi
273
info:eu-repo/semantics/conferenceObject
04 Contributo in convegno::04.01 Contributo in Atti di convegno
   NANOphotonic DEvice for Multiple therapeutic drug monitoring
   NANODEM
   FP7
   318372
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/259723
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 6
  • ???jsp.display-item.citation.isi??? ND
social impact