A high-precision exact-matching quadruple isotope dilution method (ID4MS) was employed for the quantitation of nitrate in an air-dried spinach powder Certified Reference Material (CRM). The analyte was extracted in hot water following addition of 15NO -3 internal standard. The blend was then treated with sulfamic acid to remove nitrite and with triethyloxonium tetrafluoroborate to promote aqueous conversion of nitrate into volatile EtONO2. The derivative was analyzed by headspace GC-MS with 3-min elution time. The method performance was validated with a series of tests which demonstrated adequate selectivity and ruggedness. This method supported the development of novel SPIN-1 CRM giving a modest contribution to its uncertainty (uchar = 0.85%). With respect to previous attempts, the SPIN-1 was proven stable, homogeneous (uhom = 0.44%), and suitable for spinach monitoring under EU regulations. On dried basis, the nitrate content of SPIN-1 was found to be 22.53 ± 0.43 mg/g (Uc = 1.9%, k = 2). The material was also used in an inter-laboratory study where four laboratories employed a total of ten measurement methods.
Certification of nitrate in spinach powder reference material SPIN-1 by high-precision isotope dilution GC-MS
Beatrice Campanella;Massimo Onor;Massimiliano D'Imperio;Angelo Parente;
2019
Abstract
A high-precision exact-matching quadruple isotope dilution method (ID4MS) was employed for the quantitation of nitrate in an air-dried spinach powder Certified Reference Material (CRM). The analyte was extracted in hot water following addition of 15NO -3 internal standard. The blend was then treated with sulfamic acid to remove nitrite and with triethyloxonium tetrafluoroborate to promote aqueous conversion of nitrate into volatile EtONO2. The derivative was analyzed by headspace GC-MS with 3-min elution time. The method performance was validated with a series of tests which demonstrated adequate selectivity and ruggedness. This method supported the development of novel SPIN-1 CRM giving a modest contribution to its uncertainty (uchar = 0.85%). With respect to previous attempts, the SPIN-1 was proven stable, homogeneous (uhom = 0.44%), and suitable for spinach monitoring under EU regulations. On dried basis, the nitrate content of SPIN-1 was found to be 22.53 ± 0.43 mg/g (Uc = 1.9%, k = 2). The material was also used in an inter-laboratory study where four laboratories employed a total of ten measurement methods.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.