The Single-Sample Calibration LIBS (SSC-LIBS), proposed on this journal by Rui Yuan et al., in 2019, is discussed.o We show that non-declared assumptions make the method equivalent to a conventional calibration curve approach.o As a consequence, the method cannot provide better analytical results than the ones obtained from a normal calibration curve.
Commentary: Comments on: "Accuracy improvement of quantitative analysis for major elements in laser-induced breakdown spectroscopy using single-sample calibration" by Rui Yuan et al. [Anal. Chim. Acta 1064 (2019) 11-16]
B Campanella;F Poggialini;S Legnaioli;B Cocciaro;G Lorenzetti;S Raneri;V Palleschi
2023
Abstract
The Single-Sample Calibration LIBS (SSC-LIBS), proposed on this journal by Rui Yuan et al., in 2019, is discussed.o We show that non-declared assumptions make the method equivalent to a conventional calibration curve approach.o As a consequence, the method cannot provide better analytical results than the ones obtained from a normal calibration curve.File in questo prodotto:
File | Dimensione | Formato | |
---|---|---|---|
Analytica Chimica Acta 1259 (2023) 341180.pdf
solo utenti autorizzati
Tipologia:
Versione Editoriale (PDF)
Licenza:
NON PUBBLICO - Accesso privato/ristretto
Dimensione
309.55 kB
Formato
Adobe PDF
|
309.55 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.