: Two low-cost (LC) monitoring networks, PurpleAir (instrumented by Plantower PMS5003 sensors) and AirQino (Novasense SDS011), were assessed in monitoring PM2.5 and PM10 daily concentrations in the Padana Plain (Northern Italy). A total of 19 LC stations for PM2.5 and 20 for PM10 concentrations were compared vs. regulatory-grade stations during a full "heating season" (15 October 2022-15 April 2023). Both LC sensor networks showed higher accuracy in fitting the magnitude of PM10 than PM2.5 reference observations, while lower accuracy was shown in terms of RMSE, MAE and R2. AirQino stations under-estimated both PM2.5 and PM10 reference concentrations (MB = -4.8 and -2.9 μg/m3, respectively), while PurpleAir stations over-estimated PM2.5 concentrations (MB = +5.4 μg/m3) and slightly under-estimated PM10 concentrations (MB = -0.4 μg/m3). PurpleAir stations were finer than AirQino at capturing the time variation of both PM2.5 and PM10 daily concentrations (R2 = 0.68-0.75 vs. 0.59-0.61). LC sensors from both monitoring networks failed to capture the magnitude and dynamics of the PM2.5/PM10 ratio, confirming their well-known issues in correctly discriminating the size of individual particles. These findings suggest the need for further efforts in the implementation of mass conversion algorithms within LC units to improve the tuning of PM2.5 vs. PM10 outputs.

Performance Assessment of Two Low-Cost PM2.5 and PM10 Monitoring Networks in the Padana Plain (Italy)

Giovanni Gualtieri
;
Lorenzo Brilli;Federico Carotenuto;Alice Cavaliere;Tommaso Giordano;Simone Putzolu;Carolina Vagnoli;Alessandro Zaldei;Beniamino Gioli
2024

Abstract

: Two low-cost (LC) monitoring networks, PurpleAir (instrumented by Plantower PMS5003 sensors) and AirQino (Novasense SDS011), were assessed in monitoring PM2.5 and PM10 daily concentrations in the Padana Plain (Northern Italy). A total of 19 LC stations for PM2.5 and 20 for PM10 concentrations were compared vs. regulatory-grade stations during a full "heating season" (15 October 2022-15 April 2023). Both LC sensor networks showed higher accuracy in fitting the magnitude of PM10 than PM2.5 reference observations, while lower accuracy was shown in terms of RMSE, MAE and R2. AirQino stations under-estimated both PM2.5 and PM10 reference concentrations (MB = -4.8 and -2.9 μg/m3, respectively), while PurpleAir stations over-estimated PM2.5 concentrations (MB = +5.4 μg/m3) and slightly under-estimated PM10 concentrations (MB = -0.4 μg/m3). PurpleAir stations were finer than AirQino at capturing the time variation of both PM2.5 and PM10 daily concentrations (R2 = 0.68-0.75 vs. 0.59-0.61). LC sensors from both monitoring networks failed to capture the magnitude and dynamics of the PM2.5/PM10 ratio, confirming their well-known issues in correctly discriminating the size of individual particles. These findings suggest the need for further efforts in the implementation of mass conversion algorithms within LC units to improve the tuning of PM2.5 vs. PM10 outputs.
2024
Istituto per la BioEconomia - IBE
Istituto di Scienze Polari - ISP - Sede Secondaria Bologna
AirQino
PM10
PM2.5
Padana Plain
PurpleAir
air quality
low-cost sensor
File in questo prodotto:
File Dimensione Formato  
Performance Assessment of Two Low-Cost PM2,5 and PM10 monitoring networks in the Padania Plain (Italy)_compressed (1).pdf

accesso aperto

Licenza: Creative commons
Dimensione 809.21 kB
Formato Adobe PDF
809.21 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/20.500.14243/513087
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 2
  • ???jsp.display-item.citation.isi??? ND
social impact